Standardenergy dispersive X-ray spectroscopy(EDS or EDX) using detector areas of 30 mm2on standard scanning transmission electron microscopes (STEM) can deliver element mappings with nm resolution within a few minutes. The condition is, that the detector head is small enough (in slim-line design) to get as close to the specimen for (high solid angle) and as high above the specimen (for high take-off angle) as possible. The latter helps to avoid shadowing and absorption effects.
标准干加装小30毫米2active area EDS with light element window, achieving 0.09 sr collection angle at 22° take-off angle, was used to analyze semiconductor interconnects (Fig. 1). The element distribution was mapped. The EDS data were processed quantitatively using the Cliff-Lorimer method. The calculation of theoretical Cliff-Lorimer factors in the ESPRIT software is based on the following information:
Within one series of specimens investigated under the same conditions, the Cliff-Lorimer method, using theoretically calculated factors, can be accurate within a few atomic percent relative to a chosen reference sample out of the specimen series. EDS data clearly reveal the tantalum and titanium of the Ta and TiN interconnect lining, as well as the copper and tungsten filling (Fig. 2). The titanium signal can be separated from the nitrogen signal. Si, Ta and W can be deconvolved and correctly assigned (Fig. 3).