The x-band ELEXSYS - II instrument equipped with SuperX function and a new signal processing unit (SPU).SuperX consists of a selection of high power, low noise of double gun source and high Q cavity.
The combination of these devices, the sensitivity of the CW - EPR of x-band raised an order of magnitude.We stipulated E500 for traditional measuring sensitivity and the CW - EPR spectrometer weak spacing signal-to-noise ratio of 3000:1.
Signal processing unit (SPU) is a highly integrated equipment, can satisfy the modern the CW - all acquisition mode of EPR.A total of eight signal input channel and various trigger mode, even if does not belong to standard in the field of EPR experiments can also be mastered.
Shown here is an example of SuperX system to improve sensitivity, is in the standard cavity (left) and use SuperX system (right) from the Cu squared ⁺ histidine gain spectra.In the standard is in the cavity of a faint and the spectrum of information, in the new SuperX system is a clear and good spectrum.
The front end of the run Xepr software package based on LINUX workstation software acquisition server is handling all real-time incident.
The digital resolution: 32 bits
Magnetic field uniformity: 10 mG @ 10 10 (X) (Z) х х (Y) 22 mm magnets (10 inches) or 12 mG @ 10 10 (X) (Z) х х 22 mm (9.5 inch) magnet (Y)
Maximum field strength: 14500 G (10 inches) magnets or 13000 G (9.5 inches) magnet
Absolute minimum detectable spin number (weak asphalt as the prototype) : 1.0 x 10 e9 spin number/G line width
Signal-to-noise ratio (weak asphalt as the prototype) : 3000-1
Preamplifier signal frequency range: 30 Hz to 6.5 MHz
A magnetic field the largest scope: scan - 18 kG to + 18 kG
Scanning speed: 320 microseconds / / points and five seconds
Maximum amplitude modulation: 20 G
Modulation frequency: 80 Hz to 200 kHz
Test mode: sweep phase lock, fast and transient model
Phase lock mode and harmonic detection: up to 5 order
Quick scan mode of scan rate: 200 Hz (5 ms scan 200 G)
Transient mode of sampling rate: 125 MHz
The transient mode maximum sampling points: 16000 points